0088 ACCU-CHEK is a trademark of Roche. Roche Diagnostics GmbH 68298 Mannheim, Germany www.accu-chek.com0 5241430001(02) – 2008-05Blood Glucose Met
9 Changing the battery 8710 Testing blood glucose in more than one patient – Information for healthcare professionals 9110.1 Testing blood gluc
100Symbols, problems and error messagesSymbols, problems and error messages12 In addition to test results, time and date other sym-bols and error mess
110111213141516171823456789101Symbols, problems and error messages12.1 Symbols shown in the display1 You are in the test-result memory. Either a save
102Symbols, problems and error messages3 You are in the settings for time and date.4 When an average is displayed (test-result memory): the symbol is
103Symbols, problems and error messages10 + 11 (ashing)The test strip was removed from the meter before blood or control solution was applied, for ex
104Symbols, problems and error messagesDo not use the result as a basis for therapeutic decisions. Ensure that the ambient temperature is between +10
105Symbols, problems and error messagestogether with :Averages cannot be calculated because you have not set the time and date, only test results w
106Symbols, problems and error messages The result is higher than 600 mg/dL (33.3 mmol/L). Repeat the test. Decimal point: displayed by mmol/L meters
107Symbols, problems and error messages12.2 ProblemsThe meter will not turn on. You have not removed the protective lm from the battery. Remove th
108Symbols, problems and error messagesThe meter is turned on, but the display is blank.The display is defective.Please contact your customer support
109Symbols, problems and error messages12.3 Error messagesIf an error message is displayed, turn the meter off. Depending on the situation, either by
13 Technical data 11614 Disposing of your meter 12115 System components 12216 Patents 12317 Customer support and service 12417.1 Info
110Symbols, problems and error messages You did not insert the test strip correctly or completely. Hold the test strip so the arrows printed on it a
111Symbols, problems and error messages The code chip cannot be read.Remove the code chip and re-insert it into the meter. If the error message appe
112Symbols, problems and error messages You have inserted a used test strip into the meter.Start again from the beginning with an unused test strip.
113Symbols, problems and error messages You removed the code chip while testing was in progress.Re-insert the code chip into the meter and repeat th
114Symbols, problems and error messagesA meter error has occurred. Start again from the beginning. If the error message appears again, the meter is de
115Symbols, problems and error messagesThe ambient temperature or the temperature of the meter is too low or too high for a test. Ensure that the ambi
116Technical dataTechnical data13 Meter type Accu-Chek Active (Model GC)Catalogue no./Serial no. See type plate on the back of the meterTest principl
117Technical dataMeasuring range 10–600 mg/dL (0.6–33.3 mmol/L)Sample size 1–2 µL (1 µL (microlitre) = 1 thousandth of a millilitre)Measuring time ap
118Technical dataTemperatureduring testingduring storagewithout batterywith battery+10 to +40 °C-25 to +70 °C-10 to +50 °CAtmospheric humidityduring t
119Technical dataDisplay 96-segment liquid crystal display (LCD)Interface infrared portSafety class IIILED/IRED Class 1Electromagneticcompatibilit
8756432112Getting to know your meterGetting to know your meter1 1.1 Your meter at a glance1 Display 2 S button (retrieving settings for time and date
120Technical dataPerformance analysis The performance data for the Accu-Chek Active system (Accu-Chek Active meter with Accu-Chek Active test strips)
121Disposing of your meterDisposing of your meter14 During blood glucose testing the meter itself may come into contact with blood. Used meters theref
122System componentsSystem components15 The Accu-Chek Active blood glucose monitoring system comprises the following components in addition to the met
123PatentsPatents16 US 5,366,609; US 5,424,035; US 5,463,467; US 5,889,585; US 6,055,060; US 6,906,802
124Customer support and serviceCustomer support and service17 17.1 Information service and troubleshootingFor questions on how to operate your Accu
125Customer support and servicePT. Roche Indonesia. Diagnostics Division.Gedung Artha Graha, lt.21. SCBD - Lot 25.Jl.Jendral Sudirman Kav. 52-53. Jaka
126Customer support and serviceRoche Diagnostics Asia Pacic Pte. Ltd 298, Tiong Bahru Road#16-01/06 Central PlazaSingapore 168730, SingaporeRoche Pro
127Alphabetical indexAlphabetical index18 AAlternative sites 40Apple (symbol) 40, 103Apple core (symbol) 40, 103Asterisk (symbol) 40, 102Atmos
128Alphabetical indexCode number 29, 32Coding 28Concentration table 71, 77Control solution 70stability 81Control window, test strips 43Cus
129Alphabetical indexFFlagging, result 40HHealthcare professionals 3, 91Hi (display) 45, 50, 106Hourglass (symbol) 35, 102IInfection risk 91
1413129101113Getting to know your meter9 Code chip slot10 Code chip (inserted)11 Type plate12 Battery compartment13 Battery (type CR 2032)14 Locking c
130Alphabetical indexmg/dL (symbol) 102mmol/L (symbol) 102OOverviewmeter 12symbols 101PPatents 123PC (display) 66, 106Performance analysis
131Alphabetical indexset-up (symbol) 22, 102Sources of electromagnetic interference 99Sources of interferenceelectromagnetic ~ 99light 98Stora
132Alphabetical indexTurning onchanging settings 22downloading results 66retrieving results 54testing 31Type plate 13, 17UUnit of measuremen
14Getting to know your meter1.2 The main features at a glanceFast testing The meter needs only approximately 5 seconds for each test. Easy testin
15Getting to know your meter Data transfer The meter has an infrared port enabling you to download stored test results to a computer. Applying blo
16Steps before testingSteps before testing2 2.1 Checking the contentsCheck that the contents of the package are complete. You will nd a list with th
17Steps before testing2.2 Checking the unit of measurement for blood glucoseBlood glucose results can be displayed in two differ-ent units (mg/dL or
18Steps before testing2.3 RemovingtheprotectivelmfromthebatteryThe meter is supplied with a battery inserted. The battery contacts are covered
19Steps before testing2.4 Performing a display checkPress at the same time for more than 3 secondsYou can check if all of the display elements are p
2Explanation of symbolsConsult the instructions for useCaution (refer to accompanying docu-ments). Please refer to safety-related notes in the manual
20Steps before testing2 Compare the display of your meter with the one shown here.If any of the display elements are missing or the unit shown for th
21Setting the time and dateSetting the time and date3 The time and date is not preset in new meters. You can still, perform blood glucose tests. In th
22Setting the time and datePress for more than 3 seconds3.1 OverviewTurning the meter onPress the S button and hold it down (for more than 3 seconds
23Setting the time and dateTime and date formatTime (hour, minute)YearDate (month, day)Final displayPress briey at the same timeConrmingthechange
24Setting the time and date3.2 Making settingsSetting the time and date format You can choose between two formats: 24-hour format 35 time from 0:00
25Setting the time and dateSetting the time The hours ash in the upper left corner of the display.1 Press the M button to set the hours.2 Press the
26Setting the time and dateSetting the year The year ashes in the upper right corner of the display.1 Press the M button to set the year.2 Press th
27Setting the time and dateFinal displayThe day ashes in the display.3 Press the M button to set the day.4 Press the S button, and you will see the
28CodingCoding4 A code chip is used to give the meter information about the properties of the test strips. This is why each pack of test strips contai
29Coding1 Compare the code number on the chip with the corresponding code number on the label of the test strip container.The three-digit number on t
3Intended useIntended useThe Accu‑Chek Active blood glucose meterMeter for quantitative determination of blood glucose values in fresh capillary blood
30Testing blood glucoseTesting blood glucose5 Please also consult the directions given at the end of this Chapter in the section “Notes on blood gluco
31Testing blood glucose5.2 Testing blood glucoseTurning the meter on 1 Take a test strip from the container and close the container again immediatel
32Testing blood glucosemg/dL meter mmol/L meterThe meter turns on and rst performs a default display check (approx. 2 seconds).4 Check that all the
33Testing blood glucose5 Check that this code number is the same as the code number on the label of the test strip con-tainer.If the numbers do not m
34Testing blood glucoseApplying blood with the test strip in the meterWhen the test strip and the ashing blood drop symbols are displayed, you have a
35Testing blood glucoseThe ashing hourglass symbol indicates that the test is in progress. The test is complete after approx. 5 seconds and the res
36Testing blood glucoseApplying blood with the test strip outside the meterWhen the test strip and the ashing blood drop symbols are displayed:1 Rem
37Testing blood glucose2 Prick the side of a ngertip with your lancing device.3 Encourage a drop of blood to form by gently massaging your nger to
38Testing blood glucoseThe test starts and the ashing hourglass symbol indicates that the test is in progress.The test is complete after approx. 10
39Testing blood glucoseTurning the meter off1 Remove the test strip from the meter.The meter turns itself off.You can dispose of used test strips in
4About these Instructions for useAbout these Instructions for usePlease read these Instructions for use carefully and completely before testing blood
40Testing blood glucoseFlagging test resultsYou can ag test results to describe certain events connected to the result or particular characteristics
41Testing blood glucoseWhile the test result is displayed:1 Press the S button repeatedly until the desired ag is displayed.The ags appear in the f
42Testing blood glucoseAfter choosing the desired ag, you can turn the meter off (see “Turning the meter off”), or go to the saved test results (s
43Testing blood glucose5.3 Checking the test result using the test strip control windowThe test strip itself allows you to estimate the test result a
44Testing blood glucoseAfter the testThe label on the test strip container shows blood glucose values in mg/dL und mmol/L next to each colour sample.1
45Testing blood glucose5.4 Symbols seen before, during or after testing and what they meanThe following symbols can appear on the disply before, duri
46Testing blood glucose5.5 Notes on blood glucose testing Incorrect results can cause the wrong therapy recommendation to be made and so lead to ser
47Testing blood glucose The test strips are sensitive to humidity and moisture. Only remove test strips from the con-tainer when your hands are comp
48Testing blood glucose You can touch the test eld when you apply blood, but do not rub the drop of blood into the test eld. Do not store used tes
49Testing blood glucosei Further information If you missed the code number in the display after you inserted the test strip, remove it from the met
5About these Instructions for useIn these Instructions for use you will come across three kinds of notes. Please read these very care-fully. This sym
50Testing blood glucoseMeasuring rangeThe meter measures blood glucose results in the range from 10 to 600 mg/dL (0.6–33.3 mmol/L).If the result is lo
51Testing blood glucoseImplausible results – possible sources of errorIf your meter repeatedly displays implausible results or error messages, please
52Testing blood glucoseDid you bend the test strip while inserting it into the meter?noDid you wait for the drop symbol to ash in the display before
53Using the meter as a diaryUsing the meter as a diary6 6.1 MemoryYour meter can save up to 350 test results with their time and date. You do not nee
54Using the meter as a diary6.2 Retrieving results from the memory1 With the meter turned off, press the M button.You can also go directly to the sa
55Using the meter as a diaryRetrieving older results:2 Press the S button.While you press the S button, the memory location number is displayed. When
56Using the meter as a diaryTurning the meter off again:3 Press the M button briey.You can directly switch from the saved results to the test by ins
57Using the meter as a diaryNo result in the memory Time and date not setIf there are no results saved in the meter, the display shown next to this te
58Using the meter as a diary6.3 Retrieving averagesThe meter can calculate averages from the saved results. The averages are calculated for three gro
59Using the meter as a diary1 With the meter turned off, press the M button briey.You can also briey press the M button directly after a test while
6About these Instructions for useIn these Instructions for use you will see examples of display screens. Elements that are surrounded by a halo in the
60Using the meter as a diary2 Press the M and S buttons briey at the same time.You can not only call up the averages from within the last saved resu
61Using the meter as a diaryTo go the next average:3 Press the M and S buttons briey at the same time.The averages are displayed in the order shown
62Using the meter as a diaryGo back to saved resultsorturn the meter off.Press brieyPress brieyFrom within each average you can either go back to th
63Using the meter as a diaryIf there are no results saved for the selected average, three dashes are displayed.The meter calculates the average for a
64Using the meter as a diary6.4 Downloading test resultsInfrared portYour meter has a built-in infrared port which enables wireless data downloading
65Using the meter as a diaryDepending on which type of analysis you opt for, you will require either a special software programme and/or special hardw
66Using the meter as a diary1 Read the operating instructions supplied with the software and hardware you are using. There you will nd all the infor
67Using the meter as a diary5–20 cm4 Place the meter approx. 5-20 cm away from the infrared port of the receiving end. Position the two infrared port
68Using the meter as a diaryPress brieyIf the meter was not turned off automatically by the receiving end when the download was completed, is displa
69Using the meter as a diaryIf downloading is unsuccessfulIf the results are not downloaded ( is still ashing in the display), the meter turns off a
Contents1 Getting to know your meter 121.1 Your meter at a glance ...
70Checking your meterChecking your meter7 You can check whether the meter provides correct results. To perform this performance check, a glucose contr
71Checking your meter7.1 Preparing for a performance checkFor a performance check, you need: the meter with the code chip insertedthe Accu-Chek Act
72Checking your meter7.2 Carrying out a performance check1 Read the package insert which came with the control solutions.2 Take a test strip from t
73Checking your metermg/dL meter mmol/L meterThe meter turns on and rst performs a default display check (approx. 2 seconds).5 Check that all the se
74Checking your meter6 Check that this code number is the same as the code number on the label of the test strip con-tainer.If the numbers do not mat
75Checking your meter7 Open a bottle of control solution.8 Wipe the tip of the dropper with a clean, dry paper towel to remove any dried-on residues
76Checking your meterNo agPress 4 times brieyControl (bottle)The ashing hourglass symbol indicates that the test is in progress. The test is comp
77Checking your meterFigures are examplesAfter agging the result as a performance check:12 Compare the test result with the concentration table on t
78Checking your meter13 Then remove the test strip from the meter.The meter turns itself off.You can dispose of used test strips in your house-hold w
79Checking your meter7.3 Possible sources of error during performance checksIf the result is outside the specied concentration range, repeat the per
5 Testing blood glucose 305.1 Preparing to test blood glucose ...
80Checking your meterDid you take the test strip from the container which the code chip in the meter belongs to?yesDid you perform the performance che
81Checking your meterDid you compare the test result with the concentration data that corresponds to the control solution you used?yesIs the concentra
82Cleaning the meterCleaning the meter8 You may need to clean the meter if it becomes dirty through improper use. Healthcare professionals using the
83Cleaning the meter8.1 External partsIf the case of the meter or the display is dirty:1 Wipe away any dirt using a cloth lightly mois-tened with co
84Cleaning the meter1 Pull the cover of the measuring window straight towards you.2 Wipe the cover and the test strip guide from the outside and the
85Cleaning the meter3 Carefully wipe the measuring window and its surrounding area using a lightly moistened cloth or a cotton swab. Make sure that
86Cleaning the meter6 Place the cover straight and centred onto the meter.7 Slide it closed until it audibly clicks into place.In doing so, the guid
87Changing the batteryChanging the battery9 When the battery symbol appears in the display for the rst time, the battery is almost empty. You can p
88Changing the batteryThe battery compartment is located on the back of the meter.1 Push the locking catch downwards and remove the battery compartme
89Changing the battery3 Place the new battery in the compartment, with the + symbol facing upwards.4 Re-attach the battery compartment cover so the
6 Using the meter as a diary 536.1 Memory ...
90Changing the battery5 Press the cover closed. It will CLICK into place audibly. Never throw batteries into a re. They may explode.i Dispose of u
91Testing blood glucose in more than one patient – Information for healthcare professionalsTesting blood glucose in more than one patient – Informatio
92Testing blood glucose in more than one patient – Information for healthcare professionals Patients and medical staff are potentially at risk of bec
93Testing blood glucose in more than one patient – Information for healthcare professionals10.1 Testing blood glucose1 Wear protective gloves.2 The
94Testing blood glucose in more than one patient – Information for healthcare professionals9 Encourage a drop of blood to form by gently massaging th
95Testing blood glucose in more than one patient – Information for healthcare professionals10.2 Disinfecting the meterThe following parts of the mete
96Measurement and storage conditionsMeasurement and storage conditions11 So that the meter works reliably and you obtain accu-rate results, make sure
97Measurement and storage conditions Tests cannot be performed at temperatures below +5 °C and above +45 °C. In this case the following appears in t
98Measurement and storage conditions11.2 Atmospheric humidityOnly test blood glucose when atmospheric humidity is below 85 %.Keep the meter below 93
99Measurement and storage conditions Avoid sudden changes in light conditions while testing. The ashlight of a camera, for exam-ple, can lead to inc
Comments to this Manuals