Roche Accu-Chek Aviva Technical Information Page 263

  • Download
  • Add to my manuals
  • Print
  • Page
    / 282
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 262
261
Appendix D: Meter Settings and Range Limits
mmol/L
Data Type Unit of Measurement MIN MAX Increments Default Setting
Acting Time hours:minutes 1:30 8:00 0:15 4:00
Active Insulin Units 0 99.9 0.1 N/A
bG Threshold (High) mmol/L 6.5 19.5 0.1 Hyper Warning Limit
bG Threshold (Low) mmol/L 3 5.5 0.1 Hypo Warning Limit
Carb Ratio (carbs) grams
BE
KE
CC
1
0.1
0.1
0.1
240
20
24
16
1
0.1
0.1
0.1
No entry (“--- g”)
No entry (“--- BE”)
No entry (“--- KE”)
No entry (“--- CC”)
Carb Ratio (insulin) Units 0.1 50 0.1 1
Carbohydrates grams
BE
KE
CC
0
0
0
0
240
20
24
16
1
0.1
0.1
0.1
No entry (“--- g”)
No entry (“--- BE”)
No entry (“--- KE”)
No entry (“--- CC”)
50813_00048009694_D.indb 261 7/3/12 3:14 PM
Page view 262
1 2 ... 258 259 260 261 262 263 264 265 266 267 268 ... 281 282

Comments to this Manuals

No comments