Prociency Testing
94
Applying a prociency sample
Once the meter has detected the test strip, you are
prompted to apply the prociency sample.
1 Wait until the ashing drop appears in the display
before applying the sample.
2 Apply a drop of sample to the front edge of the
test strip. Do not apply the sample to the top of the
test strip.
The sample is pulled into the test strip area by
capillary action.
Once sufcient sample has been detected, the meter
beeps and the measurement begins.