Testing with the Accu-Chek Inform II
Quality Control Testing
Should the result be outside the specified target range, proceed as
follows:
• Suspend all further testing
• Check the date on the side of the QC vials to ensure it’s
within 90 days
• Repeat the QC test
If the test continues to be outside the target range return the meter
for assessment and retrieve a back up meter