  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 78
77
5
Display Shows Possible Cause(s) Possible Solution(s)
Infrequent Pump
Communication
j
At least two weeks have passed
since the meter and pump have
communicated.
j
It is important to use Bluetooth wireless
technology communication between the
meter and pump regularly if you utilize
bolus advice.
j
Ensure the meter and pump have
Bluetooth wireless technology turned on.
j
Ensure the meter and pump are within
communication range.
j
If the low battery icon is displayed on the
meter, the meter cannot communicate
with the pump. Replace the batteries.
Invalid PIN
j
The wrong PIN was entered.
j
Select OK and re-enter the PIN shown on
the pump display.