Proficiency Testing
87
Performing a proficiency test
Overview of test procedure A proficiency test comprises the following steps:
Enter a sample ID for the proficiency sample.
Check the lot number of the test strips.
Perform the actual test with the proficiency sample.
Starting a proficiency test The following description assumes that the meter is
powered on and the main menu is displayed.
1 Touch in the Main Menu to open the Main
Menu 2 screen.
2 Touch Proficiency to start the proficiency test.
Main Menu 12:48pm
03/16/10
Review Results
Control Test
Patient Test
Main Menu 2 12:48pm
03/16/10
Maintenance
Strip Lots
Prociency
Control Lots
Beeper
Linearity
Diagnostics
Linearity Lots
Admin.