www.pasa.nhs.uk/cep Report 06024 CoaguChekXS system June 2006 Health and social care working together

Page 2

Instrument details and operation Name CoaguChek XS Manufacturer Roche Diagnostics Ltd. Bell Lane Lewes, East Sussex

Page 3 - CoaguChek XS system

Instrument details and operation Operational details Analytical procedure Prior to testing, the operator should ensure that batteries and the correct

Page 4 - Contents

Instrument details and operation Figure 4: Image of instrument and test strip, showing matching codes on LCD and test strip container Test strip cont

Page 5 - Summary

Report 06024: CoaguChek XS system Technical assessment 10 Methods For reference laboratory testing, venous blood was collected into one-tenth volume

Page 6 - Introduction

Report 06024: CoaguChek XS system Technical assessment Reference ranges The International sensitivity index (ISI) for each thromboplastin used in the

Page 7

Report 06024: CoaguChek XS system Technical assessment Table 3: Within-run imprecision for the CoaguChek XS Test 1 Test 2 Test 3 Mean INR % CV

Page 8 - Test strip

Report 06024: CoaguChek XS system Technical assessment Table 5: Between test strip lots variability using venous blood, all (n = 20) Test Strip A

Page 9 - Sample type and volume

Report 06024: CoaguChek XS system Technical assessment Table 7: INR values for oral anticoagulant samples within therapeutic range (n = 46) CoaguChe

Page 10

Report 06024: CoaguChek XS system Technical assessment Figure 5: INR for all oral anticoagulant patients. CoaguChek XS v Innovin Figure 6: INR for a

Page 11 - Operational details

Report 06024: CoaguChek XS system Technical assessment Figure 7: INR for all oral anticoagulant patients. CoaguChek XS v PT HS Plus 0123456PT HS Plus

Page 12 - Quality control

About evaluation reports The Centre for Evidence-based Purchasing provides independent and objective evaluations of medical devices available on the

Page 13 - Technical assessment

Report 06024: CoaguChek XS system Technical assessment Figure 9: INR values for OAC patients in the therapeutic range and normal subjects on the Coagu

Page 14 - Imprecision

Report 06024: CoaguChek XS system Technical assessment Stratified INR data Stratification of INR results with respect to the values obtained by the P

Page 15 - Test strip variability

Report 06024: CoaguChek XS system Technical assessment Table 8: INR subgroups for OAT samples (n = 62) CoaguChek XS Innovin PT HS Plus INR 1.00 –

Page 16 - Comparability

Report 06024: CoaguChek XS system Technical assessment Effect of haematological parameters on CoaguChek XS We assessed the effects of elevated white

Page 17

Report 06024: CoaguChek XS system Technical assessment Instrument reliability No mechanical failures were encountered during the evaluation. Only two

Page 18

Report 06024: CoaguChek XS system Technical assessment Microbiological safety The CoaguChek XS test-strips are single-use and disposable and since t

Page 19 - Mean INR

Discussion The CoaguChek XS is a small, portable device intended primarily for use by patients and non-laboratory based medical staff to generate INR

Page 20 - OAC samples Normals

Discussion values were compared with PT-HS Plus. If the average INR was greater than 4.0 units the results were excluded from this analysis. The regr

Page 21 - Stratified INR data

Conclusions This study showed that the CoaguChek XS instrument and test strips gave accurate and reproducible INR results when compared with two refer

Page 22

References CLSI E15-A. User demonstration of performance for precision and accuracy; Approved guideline (2001) Clinical and Laboratory Standards Inst

Page 23 - Carryover

Evaluation of CoaguChek XS system I Longair, C Gardiner, MA Pescott, IJ Mackie, H Cohen, SJ Machin Department of Health Evaluation Centre 6

Page 24 - User assessment

Acknowledgements We wish to thank Hilary Erwin and phlebotomy staff as well as the patients at University College Hospital for their help during this

Page 25 - Electrical safety

Appendix Electrical safety certificate Report 06024: CoaguChek XS system 28

Page 26 - Discussion

Appendix Report 06024: CoaguChek XS system 29Manufacturer’s comments

Page 27

Contents Summary ...2 Introduction...

Page 28 - Conclusions

Summary The CoaguChek XS is a portable, lightweight, point-of-care device which measures Prothrombin time (PT) and International Normalised Ratio (IN

Page 29 - References

Introduction The CoaguChek XS System is a point of care device for measuring prothrombin time (PT) and International Normalised Ratio (INR). It is

Page 30 - Acknowledgements

Introduction Figure 1: Image of the CoaguChek XS Instrument and its components LCD screen ON/OFF button Memory functions button SET button Test strip

Page 31 - Appendix

Introduction Principle components The instrument was provided in a portable case containing the CoaguChek XS meter, user manual, CoaguChek Softclix la

Page 32 - Manufacturer’s comments

Introduction Sample type and volume We initially compared fresh capillary whole blood and venous non-anticoagulated samples for equivalence using the