
Report 06024: CoaguChek XS system
Technical assessment
Instrument reliability
No mechanical failures were encountered during the evaluation. Only two error
messages from 167 tests occurred post-testing i.e. after the code numbers on
the instrument and test-strips were confirmed as matching: one error pertained
to blood application (Error 5) and the other was a measurement error (Error 6).
No quality control errors occurred..
User assessment
As with most point-of-care monitors, the CoaguChek XS only displays the INR
results to one place of decimals, whereas typical laboratory generated INR
values are recorded to two decimal places.
At present, there is no quantitative quality control material available for the
CoaguChek XS. Furthermore, as only fresh whole blood can be used, it is not
possible to test external quality assurance materials. It is recommended that
some form of external quality control as well as external quality assurance are
regularly performed on all point-of-care coagulation monitors (England et al,
1995; Fitzmaurice et al 2005). Consequently, independent quality assurance of
the CoaguChek XS would need to be performed by periodically testing a paired
capillary blood sample on a separate CoaguChek XS (or other point-of-care
monitor) maintained by the practitioner managing the oral anticoagulant care, or
using a paired venous sample tested by a laboratory based coagulometer.
Each code chip provides the instrument with lot specific calibration data and
expiry date. It was observed that a mismatch between the test strip code and
the chip code was possible. If the user places a test strip from a different lot on
the monitor, but then presses the memory button rather than replacing with
tests strips matching that of the current chip an incorrect INR value may result.
The manufacturer’s manual does emphasise the importance of matching test
strips to the current code chips but we feel that this should be further highlighted
in any protocol devised for self-testing with this instrument. Obviously a ‘fail-
safe’ electronic check would be preferable.
Each time the monitor is switched on a full display check appears showing all
symbols that may be displayed. There is a facility to call up the display after
start-up to allow checking against figures in the user manual. The manufacturer
suggests that this should be checked frequently to ensure that reading errors do
not occur.
We felt that the light weight compact size of the instrument were advantageous
when testing on patients: easily portable to where the patient was being finger-
pricked and thus testing could readily be performed in a timely manner with
minimal likelihood of blood spillages.
We also noted the low rate of test-strip failure, minimising the cost of repeat
testing.
21
Comments to this Manuals