Maintenance and Care
124
Disinfecting
Healthcare professionals must wear gloves and follow
their institution’s infection control procedures when han-
dling blood glucose testing equipment.
If your facility chooses to disinfect the Accu-Chek
Inform II system, use the following procedures.
Before using a disinfectant solution on the meter, base
unit, or code key reader, refer to the product labeling for
required contact time for disinfecting and to ensure that
the active ingredients are acceptable for use. Prepare all
disinfection solutions per the manufacturer's product
labeling. Acceptable active ingredients for disinfecting
solutions include:
0.625 % (or less) solution of sodium hypochlorite in
water (bleach)
ammonium chloride (up to 0.25 % of each
quaternary ammonium compound) with isopropyl
alcohol (up to 55 %)
0.05 % solution of a phenol compound in water
Use of disinfectants containing other active ingredients
could result in damage to the meter, base unit, or code
key reader.
Acceptable ingredients of disinfectants for the accessory
box are:
0.625 % (or less) solution of sodium hypochlorite in
water (bleach)
Use of disinfectants containing other active ingredients
could result in damage to the accessory box.
NOTICE
Do not use disinfectants containing the chemicals ether,
polyhexanide, or prepared solutions or wipes containing
a mixture of bleach and detergent on the system. Use of
disinfectants containing these chemicals could result in
damage to the system.