Appendix
141
A Appendix
Table of configuration options
This section provides an overview of all the settings avail-
able.
The two rightmost columns describe the accessibility of a
configuration parameter on the device (Setup Menu) and
via the data management system (DMS).
“Y” (Yes) means, that this parameter is available, “N” (No)
means, that this parameter is not available using the
respective configuration method.
Configuration options
DEVICE: If the meter is configured by a system
administrator using a data management system, the
configuration options on the meter may be disabled to
avoid conflicts in the settings.
DMS: Configuration options may vary depending on
available DMS settings.
Subject/Attribute Range Default Device DMS
Timer
Timer: Activity turn off 30 – 3,600 s 300 N Y
Beeper
Beeper volume 0: low
1: medium
2: high
2YY
Measurement Flow
Comments required 0: optional
1: out of range
2: required
0NY
Control lot verification 0: display only
1: yes / no confirmation
2: prompt for entry
3: scan only
1YY
Control result display 0: value (numeric)
1: PASS / FAIL
0NY
Critical range display 0: value (numeric)
1: HI / LO
0NY