Appendix
142
Critical range HI limit 10 – 600 mg/dL
0.6 – 33.3 mmol/L
600 mg/dL
33.3 mmol/L
YY
Critical range LO limit 10 – 600 mg/dL
0.6 – 33.3 mmol/L
10mg/dL
0.6 mmol/L
YY
Critical range message enabled Whether to display the
out-of-critical range
warning message (1) or
not (0)
1NY
Critical range text 0 – 100 characters “” N Y
Normal range HI limit 10 – 600 mg/dL
0.6 – 33.3 mmol/L
600 mg/dL
33.3 mmol/L
YY
Normal range LO limit 10 – 600 mg/dL
0.6 – 33.3 mmol/L
10mg/dL
0.6 mmol/L
YY
STAT tests allowed 0: no
1: yes
1YY
Number of STAT tests allowed 0 – 9 9 Y Y
Reportable range HI limit 10 – 600 mg/dL
0.6 – 33.3 mmol/L
600 mg/dL
33.3 mmol/L
YY
Reportable range LO limit 10 – 600 mg/dL
0.6 – 33.3 mmol/L
10mg/dL
0.6 mmol/L
YY
Reportable range message Whether to display a
reportable range
message (1) or not (0)
1NY
Reportable range message text 0 – 100 characters “” N Y
Strip limitations warning Configuration, whether to
display the strip
limitations warning (1) or
not (0).
0YY
Strip lot verification 0: display only
1: yes/no confirmation
2: list selection
3: scan only
1YY
Linearity lot verification 0: display only
1: yes/no confirmation
2: prompt for lot entry
1YY
Display
Low Power Mode time (only serial
numbers UU11030000)
0 - 120 sec (0 = Low
Power Mode off)
30 sec Y Y
Contrast 1 - 15 7 Y Y
Subject/Attribute Range Default Device DMS